| Subject isolation | BF (Body Floating) |
| Number of channels | 0 to 136 monopolar active electrodes + 6 bipolar auxiliary sensors |
| Output sample-rate (down-sampling available in software) | 16,384 Hz |
| Bandwidth (-3dB) | DC – 5.4 kHz, 0.002 dB ripple in passband |
| Low-pass response | 140 tap FIR filter, -65 dB @ 8 kHz (Nyquist frequency) |
| High-pass response | fully DC coupled |
| Digitalization | 24 bit SAR converter with 4x oversampling and integrated decimation filter, one converter per channel, synchronous sampling |
| Sampling skew and jitter | < 1 ns |
| Sample rate accuracy | 0.8 Hz |
| Quantization-resolution | LSB = 31.25 nV, no missing codes |
| Gain accuracy | 0.1 % |
| Anti-aliasing filter | 1st order analog filter, -3dB at 7.8 kHz |
| Total input noise (Ze < 10 kOhm), full bandwidth | 2.8 uVRMS (approx. 15 uVpk-pk), see Fig. 1, noise spectrum |
| 1/f noise (Ze < 100 kOhm), 0.1 – 10Hz | 2 uVpk-pk , see Fig. 2, pink noise plot |
| Input current noise | < 1 pArms |
| Input bias current | < 100 pA per channel |
| Input impedance | 600 MOhm @ 50 Hz (1012 Ohm // 5 pF) |
| DC offset: | < 2 mV |
| Input range | +200 mV to -200 mV |
| Intermodulation distortion, test signals: 2990 and 3010 Hz | < 0.01 % |
| Channel separation | > 100 dB |
| Common Mode Rejection Ratio | > 100 dB @ 50 Hz |
| Isolation Mode Rejection Ratio | > 160 dB @ 50 Hz |
| Power Consumption | 0.5 Watt (6 channels) to 1 Watt (142 channels) |
| Battery capacity | 3500 mAh, 3.7 V Li-ion 18650 type |
| Battery life | 10 hours (142 channels) to 20 hours (6 channels) |
| Leakage current, normal operation | < 1 uA rms. |
| Leakage current, single fault condition | < 40 uArms |
| Trigger inputs | 16 inputs on optical receiver (isolated from subject section), TTL level |
| Trigger outputs | 12 outputs on optical receiver (isolated from subject section), TTL level |
| Computer interface | USB2.0, Windows or Linux |
| Size of AD-box (width x depth x height) | 162 x 149 x 62 mm |
| Weight of AD-box, including battery | 0.8 kg |
| Environment | Indoor use: Temperature: +10°C to +40°C Humidity: 30 to 75%, non condensing Pressure: 700 hPa to 1060 hPa |
| Warranty | 3 years (1 year on electrodes) |
| Conformity | EN61326 (EMC), EN60601-1 (safety) |