AD box

New features of the ActiveThree AD box #

ActiveThree AD Box
ActiveThree AD Box
  • All channels operate at 16kHz all the time — there is no need to choose a sampling rate for the hardware. Software decimation can be used to reduce the sampling rate of the saved data.
  • Maximum of 142 channels – The new ActiveThree AD box has the capacity for 142 channels total: 136 for active biopotential electrodes and six for auxiliary sensors.  
  • Simplified upgrades – Adding auxiliary sensors to ActiveThree simply requires the purchase of a sensor. No internal electronics need to be added to support sensors. (Note: adding more biopotential measurement channels does require the AD box to be returned to us for more amplifier/converter modules to be added inside.
  • Easy hyperscanning – ActiveThree systems can be linked together for hyperscanning by means of a special hyperscanning adapter that connects AD boxes together via the Trigger port. This electrical clock synchronization replaces the optical daisy-chaining offered for ActiveTwo.  This means that hyperscanning can be added to any ActiveThree AD box.

Front panel #

ActiveThree AD Box Front Panel Showing Optical Output Connector
ActiveThree AD Box Front Panel Showing Optical Output Connector
  • Optical output port – ActiveThree uses an SC (square, plastic, shrouded) style optical fiber connector, and the port for the fiber is on the system’s front panel.

Top Panel #

ActiveThree top panel showing input connectors power button and LEDS
ActiveThree top panel showing input connectors power button and LEDS
  • High-density active electrode input connectors – A1-32, B1-32, C1-32, and D1-32 MkIII style high-density electrode input connectors.  Note that these inputs are intended for Biosemi Mk3 active electrodes only.
  • Individual active electrode input connectors – EXG 1-8 two-contact touchproof individual electrode input connectors. Note that these inputs are intended for Biosemi Mk3 active electrodes only.
  • CMS / DRL input connector. Note that this input is intended for the Biosemi Mk3 CMS / DRL electrode cable only.
  • CMS in range indicator LED – Blue LED, continuously ON when connected electrodes are within range, blinking when electrode malfunction is detected. On (lit) indicates CMS/DRL electrodes are adequately connected to the subject and no fault conditions (over/under-current) are detected. A blinking blue LED indicates EITHER that the CMS/DRL electrodes are not properly connected and gelled OR that there is some other electrode cable fault. Any single broken wire or bent pin on an electrode connector can cause a minor fault that causes the CM in range LED to go off.  See the troubleshooting section for instructions on how to identify which electrode or electrodes may be involved.
  • Power button – Press the power button momentarily to turn the power on or off. When the power indicator is illuminated in green, power is reaching the AD box from the battery.  When the power indicator is off, power has been turned off. When the power indicator is red, the battery is close to being depleted.
  • Battery indicator LED – Green LED for power ON, Red LED for low battery voltage (< 20%)

Back panel #

ActiveThree AD Box back panel showing aux ports
ActiveThree AD Box back panel showing aux ports

The back panel of the AD box has a row of six inputs labeled Aux 1-6 for use with sensors and couplers provided by Biosemi only. The following optional auxiliary sensors and couplers are available. 

  • Skin conductance (aka, GSR or galvanic skin response)
  • Respiratory effort
  • Body surface temperature
  • Pulse / plethysmograph
  • Microphone
  • Photocell
  • Optically isolated line level audio input 

NOTE: Aux inputs are on the patient-connected side of the device, so avoid connecting AC powered signal sources to these inputs.  Doing so could result in an unsafe situation and/or reduction of signal quality on other channels.

Side panel #

ActiveThree AD Box side panel showing battery compartment
ActiveThree AD Box side panel showing battery compartment

The side panel of the AD box has the battery compartment. Use only a Lithium Ion 18650 type battery. Turn the battery holder cap counterclockwise when you need to remove the cap to remove or replace the battery. Insert the battery with the positive pole outwards (protruding bump on the battery is positive). Note that the system is protected for reverse polarity. Unlocking the cap disconnects the battery from the internal electronic circuitry, so you should not remove the cap while the system is in operation.