Cautions
- Do not attempt to use the system before familiarizing yourself with the documentation. Most operator errors that result in damage to the system can be avoided by ensuring that operators are properly trained and familiar with the system documentation, especially this section and the sections covering Electrode Handling and Battery Maintenance, before use.
- Do not connect any AC powered device to the AD box. Connecting an AC powered signal source to the AD box will decrease the signal to noise ratio of other signals measured by the AD box and it can pose a safety risk. An optional Audio Optical A3 adapter to use with one of the system’s auxiliary inputs is available to connect a powered audio source (e.g. computer sound card) to the AD box.
- Do not allow active electrode tips to come into contact with any other metal. Contact between dissimilar metals can result in oxidation or reduction (redox), which may result in a loss of electrode material, discoloration of the electrode tips, and/or degradation of electrode performance (e.g. high offset potential, baseline drift, and/or increased noise).
- Do not use soap to clean gel or paste from the electrodes. Wash electrodes immediately after use with warm water to remove gel or paste residue. Soap will accelerate the loss of Chloride from the electrode pellets.
- Do not soak electrodes in any liquid for more than 10 minutes at a time. Prolonged soaking of the electrodes allows liquid to penetrate the ceramic electrode pellet, softening it, and making it more vulnerable to breakage. Soaking the electrodes in liquid for a prolonged period of time when dissimilar metals are present in the water can also result in redox, just as if the electrodes were in direct contact with a dissimilar metal.
- Do not attempt to plug connectors into the system in the wrong orientation.
Attempting to force connectors into the system in the wrong orientation can damage input connectors on the AD box, necessitating costly and time-consuming repairs. Connect 32 channel electrode cables to the AD box with the red line on the electrode cable oriented toward the labels A1, B1, C1, D1 on the left side of the AD box when facing the front panel of the AD box. Also, make sure that you can see the serial number on the side of the electrode cable connector and that you can read the model number label on the top of the electrode connector.
- Do not use alligator clips or other metal conductors to connect signal inputs to the active electrodes. Connecting alligator clips or other metal conductors to the active electrodes will result in corrosion and it will most likely result in permanent damage to the active electrodes. Use the one- and two-bucket methods described in the Troubleshooting section to short inputs or to conduct signals to the system inputs via compatible Ag/AgCl electrodes.