How are the audio cables routed into the MR scan room?

Shielded BNC cables are routed through a filter box/transformer. The supplied BNC connectors will need to be mounted within your penetration panel, this will require two 11mm holes for this purpose, the supplied RF filters will be connected on the MR side with 9 meters of cabling to connect to the headphones. Typically cables run…

Read More

What is the typical configuration for the MR Confon/BOLDfonic system?

The BOLDfonic system is comprised of the MR Confon main console, an operator microphone, a set of patient/participant insert earphones, control room monitoring speakers, along with all audio cables, connectors, and filters necessary for full installation — just supply the audio output from your PC or alternative source via a ⅛ in. audio cable or…

Read More